Recall Z-2146-2016 ยท KPA Slide Stainer, Automated

Hardy Diagnostics

QuickSlide MGS-80, Automated Mutli-Slide Gram Stainer/GramPRO-80. Systematically stains, rinses, decolorizes, and counter-stains the provided biological sample and slide.

DateJul 8, 2016 (initiated May 24, 2016)
StatusTerminated
ReasonDiscrepant culture and gram stain results, which could potentially lead to misdiagnosis or improper treatment of a patient.
Root causeUnder Investigation by firm
ActionThe firm began notifying consignees via phone on 05/23/16. The firm also sent out Urgent Notification letters, dated June 15, 2016. The firm's phone script instructs customers to decrease the number of rinses on the slide and increasing the number of rinses in the chamber for better outcomes and performance with the instrument. Reducing the rinses on the slide may prevent bacteria from coming off the slide, and adding rinses to the empty chamber will help ensure that the chamber is clean before the next slide is introduced. The firm's customer notification letter recommended an immediate u
Quantity25
DistributionU.S. Distribution to the following states: CA, TX, NE, OK, LA, NC, PA, AZ, NY, PA, VA, NM, MI, and WV.