Recall Z-2002-2016 ยท DWE Tubing, Pump, Cardiopulmonary Bypass

Datascope Corporation

Small Patient ECC Pack, REF BEQ-TOP 24100, Product Usage: The HLM Tubing Sets with Bioline Coating are for single use only. They may be sold sterile, non-sterile and bulk packed. Custom tubing sets that are sold sterile are not to be re-sterilized by the user. In open heart surgery the HLM Tubing Sets with Bioline Coating are used in combination with the heart-lung machine for the oxygenation of b

DateMay 9, 2016
StatusTerminated
ReasonCustom Tubing Set BEQ-TOP-24100 (Small Patient ECC Pack) was identified as possessing a level of endotoxin that exceeded the regulatory limits.
Root causeUnder Investigation by firm
ActionMaquet sent via express mail and hand delivered a recall letter and response form dated May 9, 2016 to their sole customer.
Quantity17 units
DistributionUS in the state of GA
FDA recordOpen on fda.gov