ClearanceBrief
Recall Z-2471-2016 ยท NHA Abutment, Implant, Dental, Endosseous

Biomet 3i, LLC

Implant EP Twist Lock TM Coping Dental implants

DateAug 12, 2016 (initiated Nov 25, 2015)
StatusTerminated
ReasonPouches may not have been sealed during packing.
Root causePackaging
ActionBiomet sent an Urgent Device Recall Notice to all affected consignees the week of November 30, 2015, communicating the issue and customer responsibilities. International consignees were notified via traceable courier or email week of November 30, 2015. Customers were instructed to return affected product by calling BIOMET at 1-800-342-5454 to obtain a RMA number to return products. Customers were also asked to complete the Business Reply Form and fax to 561-514-6316. For questions regarding this recall call 561-776-6700.
Quantity1, 648,273 devices (all products subject to this recall)
DistributionWorldwide Distribution Nationwide andHawaii, Guam, Puerto Rico and Virgin Islands (U.S.) . FOREIGN CONSIGNEES: Albania, Andorra, Argentina, Australia, Austria, Barbados, Belgium, Bolivia, Brazil, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Denmark, Dominican Republic, France, German
FDA recordOpen on fda.gov

510(k) clearances named in this recall