Recall Z-2122-2016 ยท EOX Esophagoscope (Flexible Or Rigid)

Covidien, LLC

Covidien Cytosponge Cell Collection Kit. Contains Cytosponge Cell Collection Device, Specimen Jar with preservative. Item code CYTO-KIT-R Gastroenterology: Cytosponge Cell Collection Device is indicated for use in the collection and retrieval of surface cells in the esophagus. Currently the device is only available for clinical studies.

DateJun 6, 2016
StatusTerminated
ReasonMedtronic is recalling all lots of the Covidien Cytosponge Cell Collection device after two reports of the device detaching from the removal string during withdrawal from the patient's esophagus.
Root causeProcess control
ActionThe firm, Medtronic, sent an "URGENT MEDICAL DEVICE RECALL" letter dated June 6, 2016 to affected customers. The letter described the product, problem and actions to be taken. The customers were instructed to quarantine and discontinue use; return affected product to Medtronic/Covidien, Quality Department, 540 Oakmead Parkway, Sunnyvale CA 94085; complete and return the RECALLED PRODUCT RETURN FORM by email to Customer Service feedback.customerservice@Covidien or fax to (800)-895-6140 if product purchased directly from Medtronic; or by email to Quality Compliance MITGFCA@Covidien.com or fax to
Quantity620 total units (260 US; 360 OUS)
DistributionWorldwide Distribution-US Distribution to TN, NC, IL, CA, MN, OR and NY; and countries of: United Kingdom and Iran.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K152794Cytosponge Cell Collection Device
Covidien, LLC, Apr 12, 2016