Recall Z-2867-2016 ยท MWI Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)

Invivo Corporation

Philips Expression Information Portal (IP5) Model 865471. The IP5 is intended to be used as a secondary display control unit for the Expression MRI Patient Monitoring System and the Expression MR200 MRI Patient Monitoring System.

DateSep 14, 2014
StatusTerminated
ReasonFrozen Display Numerics and Disabled Menu Keys after extended run time. This customer notification was sent August 22, 2014.
Root causeSoftware in the Use Environment
ActionA recall notification letter was sent to hospitals who purchased the recalled product via UPS March 2014. The letter instructed users to power cycle the IP5 unit every day. This is done by pressing the power button on the side of the unit. Customers were instructed to communicate the recall to all those subject and to attach an addendum to the IP5 Instructions for Use. The firm advised customers to call their Philips representative, option 1, then option 2, if they had any questions or concerns.
Quantity639 units
DistributionWorldwide distribution. US nationwide, Australia, Austria, Denmark, France, Germany, Iceland, Ireland, Italy, Netherlands, Norway, Poland, Saudi Arabia, Singapore, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, and England
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121424EXPRESSION INFORMATION PORTAL
Invivo Corporation, Jun 13, 2012