Recall Z-2087-2016 ยท LXG Equipment, Laboratory, General Purpose, Labeled Or Promoted For A Specific Medical Use

Siemens Healthcare Diagnostics Inc.

Hettich Centrifuge Modules: StreamLAB Centrifuge Module (10372372 or 10444807 or 10706836) Aptio Centrifuge Module (10703032 or 10713761) FlexLab High Throughput Centrifuge Module (11171747) FlexLab and Aptio Centrifuge Module (10628157 or 11266581): Used to separate substances according to their densities.

DateMay 11, 2016
StatusTerminated
ReasonPremature bucket failures
Root causeUnder Investigation by firm
ActionSiemens mailed to customers an Urgent Medical Device Correction (UMDC#LAI16-02.A.US) Urgent Field Safety Notification letter informing customers of the problem with recommendations to follow the existing maintenance instructions in the product operator manuals as well as a new requirement to spin at cycle speeds not exceeding 3500 rpm.
Quantity487
DistributionNationwide: AL, AR, AZ, CA, CO, CT, FL, GA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA, WI, WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.