ClearanceBrief
Recall Z-2085-2016 ยท HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)

Medtronic Sofamor Danek Usa, Inc - Dallas Distribution

Saw Sagittal Catalog Number: ES 300; Saw Sagittal Refurbished Catalog Number: ES300-R; Saw Saggital, Finger Catalog Number: ES310; Saw Saggital, Finger Refurbished Catalog Number: ES310-R; Used to dissect bone in surgical applications.

DateJun 28, 2016 (initiated May 13, 2016)
StatusTerminated
ReasonMedtronic Powered Surgical Solutions testing of the Sagittal Saw Rocker Arm Seal identified a potential for leakage of fluid into the motor during use.
Root causeDevice Design
ActionRecall notification letters were sent to consignees on 05/13/2016.
Quantity368 devices
DistributionWorldwide Distribution-Including the states of AL, AR, AZ, CA, CT, FL, GA, IA, ID, IL, IN, MA, MD, NC, NH, NY, SC, TN, TX, and UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081475ELECTRIC DRILL SYSTEM [XPS 4000 SYSTEM, MIDAS REX LEGEND EHS SYSTEM, INTEGRATED POWER CONSOLE (IPC)]
Medtronic Xomed, Inc., Oct 17, 2008