Recall Z-1929-2016 ยท IYE Accelerator, Linear, Medical

Brainlab AG

ExacTrac Vero is a Patient Positioning System for Radiation therapy.

DateMay 4, 2016
StatusTerminated
ReasonPotentially incorrect positioning when using Implanted Marker Detection with Brainlab ExacTrac Vero 3.5
Root causeDevice Design
ActionA FIELD SAFETY NOTICE / PRODUCT NOTIFICATION was sent to customers on May 4, 2016 and distributed via email. An active reply by the consignees will be followed, if necessary via further different communication media. As of now for new installations - the Product Notification letter is handed over to the customer during installation (before clinical use) by the according Brainlab Service representative, - a signature is required by the customer that the contained User Corrective Action will be implemented before clinical use by the facility (to be adhered to by the users). Brainlab Instructio
Quantity27 Systems
DistributionDistributed in the states of Florida, New York, Texas and Ohio, and in the countries of Belgium, France, Germany, Italy, Japan and South Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122451EXACTRAC VERO
Brainlab AG, Sep 28, 2012