Recall Z-2248-2016 ยท JEA Table, Surgical With Orthopedic Accessories, Ac-Powered

Steris Corporation

OT1000 Series Orthopedic Surgical Tables

DateMay 23, 2016
StatusTerminated
ReasonSTERIS has identified that the pivot pin assembly located on the table top may become loose preventing the table from articulating properly.
Root causeOther
ActionRecall Notification Letters to the affected Consignees was sent on May 23, 2016 via FedEx with tracking numbers for delivery confirmation. STERIS Field Service Technicians will visit each domestic Customer location to perform the inspection and reinstallation of the pivot pin assemblies. For international direct accounts who have further distributed the surgical table, Steris asked to identify the Customers, and to notify them of this product field correction, and that not to operate the tables until the inspection has been done of the pivot pin assembly on all affected tables.
Quantity148
DistributionAL;AK, AZ; AR; CA; CO; CT; DC; FL; GA; IL; IN; IA; NY; NJ; NC; OH; OK; OR; PA; PR; SC; TX; UT and VA. Foreign accounts: Canada; France; Korea; Panama; Qatar and Thailand.