Recall Z-1999-2016 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA Inc.

BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10528956 Product Usage: The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

DateJun 15, 2016 (initiated Apr 25, 2016)
StatusTerminated
ReasonThe bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.
Root causeProcess design
ActionSiemens sent a Customer Safety Advisory Letter to all affected customers. The letter identified the affected product, problem and actions to be taken. Siemens service organization will be contacting customers to schedule the update as per the attached Update Instruction and will be provided the replacement bracket to resolve the issue. For questions contact your Siemens representative.
Quantity1 unit
DistributionWorldwide Distribution - US Nationwide and the countries of China and Malaysia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123737BIOGRAPH MCT FAMILY
Siemens Medical Solutions USA, Inc., Jan 29, 2013