ClearanceBrief
Recall Z-2030-2016 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA Inc.

Biograph mCT-S(40) 3R, MATERIAL NUMBER 10248668 The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.

DateJun 21, 2016 (initiated Apr 22, 2016)
StatusTerminated
ReasonSiemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.
Root causeNonconforming Material/Component
ActionSiemens sent a Customer Safety Advisory Letter on April 22, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Siemens representative at the contact numbers provided in the letter.
Quantity47 units
DistributionWorldwide Distribution - US Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141971BIOGRAPH TRUE PET/CT FAMILY SOFTWARE
Siemens Medical Solutions USA, Inc., Sep 12, 2014