ClearanceBrief
Recall Z-1986-2016 ยท LJS Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days

Navilyst Medical, Inc., an AngioDyamics Company

BioFlo Hybrid PICC with ENDEXO and PASV Valve Technology, Catheter Kit, UPN H965459410, Rx ONLY The Xcela PICC with PASV, Xcela Hybrid PICC with PASV, BioFlo PICC with PASV and BioFlo Hybrid PICC with PASV are indicated for short or long-term peripheral access to the central venous system for intravenous therapy, including but not limited to, the administration of fluids, medications and nutrients

DateMar 14, 2016
StatusTerminated
ReasonPICC catheters that contain valve(s) manufactured prior to July 15, 2014 lack a manufacturing specification change that reduces the incidence of hemolysis during blood sampling through the PICC catheter.
Root causeDevice Design
ActionNavilyst Medical distributed Urgent Voluntary Medical Device Recall notification letters on March 14, 2016, and Reply Tracking forms to their customers via courier service. Customers were asked to immediately remove any affected (recalled) product from their inventory (whether in labs, central supply, shipping and receiving or any other location). Segregate this product in a secure location for return to Navilyst Medical. Immediately forward a copy of the recall notice to all sites to which you have distributed affected product. If affected product is located in your institution, please c
Quantity10 units
DistributionWorldwide Distribution US (Nationwide) Internationally to AU, GR, SE, SA, IN, CA, GB, BR, BE, NL, ES, KR, HK, AE, IT and ES.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021704VAXCEL PICC WITH PASV VALVE TECHNOLOGY
Catheter Innovations, Inc., Jun 6, 2002