Recall Z-1837-2016 ยท NAY System, Surgical, Computer Controlled Instrument
Intuitive Surgical, Inc.
SEAL,IS4000 PORTS,12MM&STAPLER,BOX of 10; Model numbers: 470380-04, 470380-05, 470380-06; Part number:470380. The da Vinci Xi Surgical System cannula seal is used with the cannula accessory and provides sterile disposable cap for each cannula. When attached to the cannula, the seal maintains insufflation when the cannula is empty and when the instrument is inserted during a surgical procedure.
| Date | May 27, 2016 (initiated May 9, 2016) |
| Status | Terminated |
| Reason | Potential for particulates in insufflation stopcocks on the da Vinci Xi Surgical System cannula seals. The particulates could be introduced into the cannula lumen when insufflation is connected during a surgical procedure. |
| Root cause | Manufacturing material removal |
| Action | Upon discovery of the issue, Intuitive sent Urgent Product Safety Notices to all customers by e-mail to initially alert them of the situation. This will be followed by a formal letter that is scheduled to arrive on May 23, 2016. Both notification request that users Do NOT use the da Vinci Xi 5 mm8 mm Universal Cannula Seal or the da Vinci Xi 12 mm & Stapler Universal Cannula Seal for insufflation, or open the stopcock during intracardiac procedures. |
| Quantity | 46,191 boxes total - both sizes |
| Distribution | Worldwide distribution. US, Australia, Austria, Belgium, Chile, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Monaco, Netherlands, Norway, Puerto Rico, Portugal, Qatar, Romania, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzer |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K140553 | ENDOWRIST STAPLER 45 AND STAPLER 45 RELOADS Intuitive Surgical, Inc., Jul 25, 2014 |