Recall Z-0072-2017 ยท JLW Radioimmunoassay, Thyroid-Stimulating Hormone

Siemens Healthcare Diagnostics, Inc.

ADVIA Centaur XPT Immunoassay System

DateApr 28, 2016
StatusTerminated
ReasonThe ADVIA Centaur XPT default setting for the Daily Maintenance Task (Daily Cleaning Procedure) frequency may have the Daily Maintenance Task frequency set to, As needed instead of Daily. Not performing the daily maintenance may impact any assay and can be detected through monitoring of quality control and calibration results. Systems running in a language other than English are not impacted.
Root causeUnder Investigation by firm
ActionSiemens sent an Urgent Medical Device Correction letter dated April 28, 2016, to all affected customers. The letter advised customers not to release results generated by the system if quality control results are not within acceptable ranges. Customers were advised to perform the following steps to check the Setting for the Daily Maintenance Task frequency on your instrument: *Open the Maintenance Tab, * Select Daily Maintenance task and * Verify that the Frequency Column is set to "Daily at..". The Daily Maintenance Task frequency setting can only be reset by a Siemens Customer Service en
Quantity292 units Total (3 domestically & 289 internationally)
DistributionWorldwide Distribution - US including MD, MN, & MS; **Internationally** AE, AT, AU, BE, BR, CA, CN, DE, DK, EG, ES, FI, FR, GB, GP, HU, IN, IT, JO, JP, KR, MY, NO, NZ, PH, PT, QA, SA, SE, SG, TR & VN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141999ADVIA CENTAUR TSH