Recall Z-2083-2016 ยท OWB Interventional Fluoroscopic X-Ray System

Medtronic Navigation, Inc.-Littleton

Medtronic Navigation O-arm 02 surgical x-ray imaging system

DateMar 17, 2016
StatusTerminated
ReasonReview of on-system electronic product labeling and Instr uctions for Use in response to an internal management concern uncovered gaps in compliance with the requirements for information to be provided to users in 21CFR part 1 020.30(h).
Root causeRadiation Control for Health and Safety Act
ActionMedtronic planned action to bring device into Compliance: 1. You will contact customers and provide an Errata sheet to them updating the missing and incorrect information. 2. Medtronic field service personnel will perform radiation output measurements via validated procedures to attain compliance with 21CRF part 1020.32(k)(6). 3. You will update the systems software-based instructions for use by a software update which includes the information supplied by the Errata sheet. For questions call (800)-595-9709.
QuantityUS - 63
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151000O-Arm O2 Imaging System
Medtronic, Inc., Aug 6, 2015