Recall Z-1957-2016 ยท MUU System, Suction, Lipoplasty

Miami Fat Supply, Inc

The Red Head 2 Reusable Fat Harvesting Canister, Product # RH-2, Non-Sterile. Intended for use in aspirating subcutaneous fatty tissue in patients desiring aesthetic body contouring and autologous fat collection.

DateApr 21, 2016
StatusTerminated
ReasonDuring an FDA inspection it was found that the products are marketed without a cleared 510k, nor is approved for multiple use.
Root causeNo Marketing Application
ActionCustomers were sent an Urgent: Medical Device Recall/Correction letter, dated 04/19/2016, on 04/21/2016 that served as a notification of the Red Head 2 voluntary recall. Customers were informed that there were previous complaints of compromised product integrity and performance following sterilization. Lids are subject to fracture and fail after steam sterilization. Customers were advised to destroy used product and to ship unused product back to the manufacturer. Customers were to return an acknowledgement form to confirm their notification of the voluntary recall.
Quantity443 units
DistributionWorldwide Distribution -- US, including the states of AL, AZ, CA, CO, FL, GA, IL, MD, MI, MO, NM, NY, OH, TX, UT, and VA; and, the country of Dominican Republic.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.