Recall Z-2082-2016 ยท OXO Image-Intensified Fluoroscopic X-Ray System, Mobile

Medtronic Navigation, Inc.-Littleton

Medtronic 0-arm Imaging mobile x-ray system. Model Numbers - B1-700-00027, B1-700-00028

DateMar 17, 2016
StatusTerminated
ReasonIt has been discovered that replacement x-ray generator and system motion batteries could become damaged during shipment, resulting in unacceptable risk of failure.
Root causeRadiation Control for Health and Safety Act
ActionMedtronic's planned action to bring device into Compliance. 1. You will provide an Errata sheet to customers updating the missing and incorrect information. 2. Field Service Engineers (FSEs) will perform radiation output measurements via validated procedures to attain compliance with 21CFR part 1020.32(k)(6). 3. You will inspect and test X-ray generator and system motion batteries for damage, in accordance with the plan submitted to CDRH. Damaged batteries will be replaced. 4. You will update the systems software-based instructions for use by a software update which includes the info
QuantityUS - 581
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092564O-ARM IMAGING SYSTEM
Medtronic Navigation, Inc. (Littleton), Jun 17, 2010