Recall Z-1917-2016 ยท MBH Prosthesis, Knee, Patello/Femorotibial, Semi-Constrained, Uncemented, Porous, Coated, Polymer/Metal/Polymer

Zimmer Biomet, Inc.

Persona Trabecular Metal Tibia; Persona Trabecular Metal Tibial Plate Instruments and Modular Brackets; Used to facilitate the implantation of joint replacement implants.

DateJun 7, 2016 (initiated May 11, 2016)
StatusTerminated
ReasonZimmer Inc. is initiating a recall of specific Persona" Trabecular Metal" Tibial Plate Instruments and Modular Brackets, due to incidents of radiolucent lines and loosening in implants.
Root causeComponent design/selection
ActionZimmer Inc. is initiating a voluntary recall of Persona" Trabecular Metal" Tibia / Tibial Plate Instruments and Modular Brackets following an increase in complaints of radiolucent lines and loosening, and earlier recall of the Persona" Trabecular Metal" Tibial Plates . All sizes and lots of the affected devices are being removed from distribution. Recall notices were issued to affected distributors, hospitals, and surgeons on 05/11/2015 via mail. Customers may contact the following with any questions: 1-877-946- 2761 between 8:00 am and 5:00pm EST. Customers are asked to do the following: 1.
Quantity10,256
DistributionWorldwide Distribution -- US, including the states of AK, AL, AZ, CA, FL, GA, HI, ID, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, and WI; Foreign: Australia, New Zealand, Korea, Austria, Belgium, Germany, Italy, France, Luxembourg, Netherlands, South Africa,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121771ZIMMER PERSONA THE PERSONALIZED KNEE SYSTEM
Zimmer, Inc., Nov 7, 2012