Recall Z-2064-2016 ยท FRN Pump, Infusion

CareFusion 303, Inc.

Alaris System PC Unit Model 8015 with software versions 9.17 and 9.19; Central programming, monitoring and power supply component for the Alaris System.

DateMay 12, 2016
StatusTerminated
ReasonA patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110.
Root causeSoftware design
ActionAn Urgent Medical Device Recall Letter was sent to customers on 5/31/16 to inform them that BD had identified an issue with the Alaris System PC unit model 8015 with software versions 9.17 and 9.19. Firm has received reports that a patient weight can be populated incorrectly under certain conditions when using the RESTORE feature to restore infusions running on the Alaris LVP module model 8100 and the Alaris Syringe module model 8110. The letter informs the customers of the potential risk and the required action for users. Customers are informed that BD will contact them within 60 days to in
Quantity69,693 units
DistributionDistributed US (nationwide) and in Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051641ALARIS SYSTEM PC UNIT, MODEL 8001
Cardinal Health, Alaris Products, Oct 20, 2005