Recall Z-2216-2016 ยท OCC Respiratory Virus Panel Nucleic Acid Assay System

Roche Molecular Systems, Inc.

cobas Liat Analyzer

DateMay 9, 2016
StatusTerminated
ReasonDevice was released with default pcal/gain settings for photometers instead of the pcal/gain settings that are required for release. The default settings can potentially lead to the generation of erroneous or invalid results.
Root causeUnder Investigation by firm
ActionCustomers were notified of the recall via phone and email sent on 5/9/2016. They were instructed to take the following actions: Immediately discontinue use of the affected instruments until they have been reworked by a Roche Field Engineering Specialist. A Roche representative will contact you to schedule this service. Complete the attached fax form and fax it to 1-317-521-4815. File this Urgent Medical Device Correction (UMDC) for future reference. Please contact the Roche Support Network Customer Support Center 1-800-800-5973 if you have questions.
Quantity2
DistributionDistributed in NC and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111387LIAT (TM) INFLUENZA A / B ASSAY, LIAT (TM) ANALYZER
Iquum, Inc., Aug 4, 2011