Recall Z-1705-2016 ยท IYE Accelerator, Linear, Medical

Elekta, Inc.

iGUIDE System, for patient positioning, with assistance of a 30 Tracking System in a radiotherapy environment.

DateMay 23, 2016 (initiated May 12, 2016)
StatusTerminated
ReasonIf by mistake the initial pre-treatment imaging was performed BEFORE the HexaPOD was moved to the DRIVE (*START) position, iGUIDE offers the possibility to branch into a specific catch-up workflow, which was implemented to use the initial scan data.
Root causeSoftware design
ActionElekta sent an Important Field Safety Notice that was distributed to all affected customers on May 12, 2016. The notice informs end users about the problem and the potential clinical impact. It also gives a recommendation how the risk can be mitigated. The notice also advises the customer that a patch will be released that will introduce a patch that avoids an inaccurate position and thus eliminates an unsafe situation. Corrective Action #2: Permanent Solution Software Upgrade A software patch will be released that will correct the behavior of the iGUIDE software. The target release date f
Quantity20
DistributionIL, LA, Austria, Australia, Botswana, Germany, Denmark, France, India, Japan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072079IGUIDE SYSTEM
Medical Intelligence Medizintechnik-GmbH-An Elekta, Aug 14, 2007