Recall Z-1696-2016 ยท KPR System, X-Ray, Stationary

Philips Healthcare

Philips Healthcare DuraDiagnost stationary X-ray system

DateJun 3, 2016 (initiated Nov 10, 2015)
StatusTerminated
ReasonThe detector may signal that it is ready for acquisition when it actually is not, resulting in failure to properly acquire the X-ray image.
Root causeRadiation Control for Health and Safety Act
ActionPhilips Healthcare plans to correct the defect at no cost to customers by implementing a field correction involving the installation of new software. 1. You will contact customers and initiate a software update to correct the defect. 2. You will install the software and perform testing to ensure the software update was effective. 3. The customer notification letter which includes a statement that you will without charge, remedy the defect or bring the product into compliance. For further questions, please call 1-800- 722-9377.
Quantity5
DistributionUSA (nationwide) Distribution to the states of : ( DuraDiagnost) to the states of : OH, CA, WA and NE. (DigitalDiagnost) to the states of : AK, AZ, AL, CA, CT, CO, DE, FL, GA, HI, ID, KY, LA, MA, MN, MD, MS, ME, MO, NE, NY, ND, OH and PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141381DURADIAGNOST
Philips Healthcare (Suzhou) Co., Ltd., Jun 12, 2014