Recall Z-1841-2016 ยท GDZ Handle, Scalpel

Integra York PA, Inc

Miltex Premium Grade Self-Lock Chisel Blade Handle, Catalogue No. 4-405, manual surgical instrument.

DateMay 31, 2016 (initiated Mar 11, 2016)
StatusTerminated
ReasonOne lot of self-lock chisel blade handles have a wider slot that holds the chisel blade that may result in the inability to firmly secure the blade in the handle.
Root causeNonconforming Material/Component
ActionIntegra mailed out an Urgent: Voluntary Medical Device Recall letter to customers informing them to evaluate their inventory for the affected product. If product was shipped to customers, they were asked to identify their customers and to stop shipping immediately. Integra requested that customers complete the attached form that was sent with the notification indicating whether or not they had the affected product on hand and customer service will contact them to provide a Return Authorization number for the product's return. Customers could reach Customer Service at 866-854-8300.
Quantity678 Chisel Blade Holders
DistributionWorldwide distribution. US nationwide, Mexico, and Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.