Recall Z-1989-2016 ยท NZW Heparin, Vascular Access Flush

Excelsior Medical Corp

Excelsior Disposable Syringe Heparin Lock Flush, USP syringe (5 mL Fill in 10 mL syringe). For flushing of IV catheters and IV tubing only.

DateJun 14, 2016 (initiated Apr 28, 2016)
StatusTerminated
ReasonCertain lots of Heparin Lock Flush, USP Syringes (5 ml Fill in 10 ml Syringe) failed Anti-Factor IIa assay for potency during testing of retention samples for heparin potency.
Root causeNonconforming Material/Component
ActionExcelsior Medical Inc. sent a recall letter on 4/29/2016 via certified mail. Customers are instructed to discontinue using the product subject to this recall. Customers are requested to complete the customer reply form and fax it (866) 807-3714, so arrangements can be made to return the product to Excelsior.
Quantity184,500
DistributionUS Nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K023740SYREX PRE-FILLED SYRINGE
Excelsior Medical Corp., May 13, 2003