ClearanceBrief
Recall Z-1814-2016 ยท FMF Syringe, Piston

Biomet, Inc.

PerFuse Decompression Instrument PN: 800-0541, Lot:100650. Designed to access the femoral head for decompression.

DateMay 25, 2016 (initiated Apr 11, 2016)
StatusTerminated
ReasonThe Trocar and Plunger Assemblies are missing from the instrument, which could result in a delay in surgery greater than 30 minutes.
Root causeProcess control
ActionZimmer Biomet has initiated a voluntary recall of the PerFuse Decompression Instrument, following an investigation which identified that the Trocar and Plunger Assemblies are missing from the instrument. The firm contacted customers via telephone and e-mail on March 28, 2016; and official recall notices were mailed via FedEx on 4/11/2016. Customers were instructed to quarantine recalled items, Complete the Certificate of Acknowledgment and email a copy to CPWARFieldAction@zimmerbiomet.com; Adverse reactions or quality problems experienced with the use of this product may be reported to the
Quantity22
DistributionDistributed in the US states of CA, GA, MN, and OH, and in the Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141762PERFUSE PERCUTANEOUS DECOMPRESSION SYSTEM
Biomet, Inc., Aug 27, 2014