Recall Z-1829-2016 ยท KWP Appliance, Fixation, Spinal Interlaminal

RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)

Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve. Spinal fixation device instrument.

DateMay 26, 2016 (initiated Apr 22, 2016)
StatusTerminated
ReasonPotential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip, resulting in possible patient injury.
Root causeProcess design
ActionRTI sent an Urgent Voluntary Product Notification dated April 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. RTI Surgical is conducting a voluntary recall of the Streamline OCT Occipito-Cervico-Thoracic System, Crosslink Counter Torque Sleeve due to the potential for a product defect where the cannula may be obstructed. This obstruction could prevent a driver from going through, which increases the potential for the driver to slip resulting in possible patient injury. For further questions, please call (9
Quantity17 devices
DistributionUS nationwide distribution in CA, FL, IN, MI, MO, NC, NY, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150254Streamline OCT Occipito-Cervico-Thoracic System
Pioneer Surgical Technology, Inc. (Dba Rti Surgical, Inc.), Apr 28, 2015