Recall Z-1684-2016 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Zimmer Manufacturing B.V.

Trilogy self-tapping bone screw Bone screw 4.5x25 selftap 00625004525 Bone screw 4.5x50 selftap 00625004550 Bone screw 6.5x15 selftap 00625006515 Bone screw 6.5x20 selftap 00625006520 Bone screw 6.5x25 selftap 00625006525 Bone screw 6.5x30 selftap 00625006530

DateMay 16, 2016 (initiated May 4, 2016)
StatusTerminated
ReasonA small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
Root causePackaging change control
ActionAll distributors will be notified via electronic mail. Distributors with affected lots will also be notified via courier. Hospitals risk managers and surgeons will be notified via recall notification letter, dated May 4, 2016.
Quantity3893
DistributionWorldwide distribution. US nationwide, Korea, Japan, Argentina, Brazil, Singapore, Taiwan, China, El Salvador, Australia, India, Hong Kong, Chile, Germany, Italy, France, Spain, UK, Sweden, Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K934765TRILOGY ACETABULAR SYSTEM
Zimmer, Inc., Apr 29, 1994