Recall Z-1681-2016 ยท HSB Rod, Fixation, Intramedullary And Accessories

Zimmer Manufacturing B.V.

ITST intertrochanteric/subtrochanteric fixation 4.5 mm diameter cortical screw, 3.5 mm hex head, 57 mm length 00225305745

DateMay 16, 2016 (initiated May 4, 2016)
StatusTerminated
ReasonA small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
Root causePackaging change control
ActionAll distributors will be notified via electronic mail. Distributors with affected lots will also be notified via courier. Hospitals risk managers and surgeons will be notified via recall notification letter, dated May 4, 2016.
Quantity50
DistributionWorldwide distribution. US nationwide, Korea, Japan, Argentina, Brazil, Singapore, Taiwan, China, El Salvador, Australia, India, Hong Kong, Chile, Germany, Italy, France, Spain, UK, Sweden, Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143321ITST Intertrochanteric/Subtrochanteric Fixation System
Zimmer, Inc., May 29, 2015