Recall Z-1679-2016 ยท HWC Screw, Fixation, Bone

Zimmer Manufacturing B.V.

Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Fully Threaded, 5.0 mm Diameter, 20 mm Length 00114205020 Mini Magna-Fx Cannulated Screw Fixation System Small Cannulated Bone Screw, Fully Threaded, 5.0 mm Diameter, 36 mm Length 00114205036 Mini MagnaFx Cannulated Screw Fixation System Small Cannulated Bone Screw, Partially Threaded, 5.0 mm Diameter, 24 mm Length 00114205

DateMay 16, 2016 (initiated May 4, 2016)
StatusTerminated
ReasonA small hole may be present in the inner tray of specific lots of sterile Hip and Trauma Implant Screws.
Root causePackaging change control
ActionAll distributors will be notified via electronic mail. Distributors with affected lots will also be notified via courier. Hospitals risk managers and surgeons will be notified via recall notification letter, dated May 4, 2016.
Quantity751
DistributionWorldwide distribution. US nationwide, Korea, Japan, Argentina, Brazil, Singapore, Taiwan, China, El Salvador, Australia, India, Hong Kong, Chile, Germany, Italy, France, Spain, UK, Sweden, Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142442MAGNA-FX AND MINI MAGNA-FX CANNULATED SCREW FIXATION SYSTEM
Zimmer, Inc., Oct 3, 2014