Recall Z-2623-2016 ยท DQY Catheter, Percutaneous

Cook Inc.

White Lumax Guiding Coaxial Catheter Catheter, Percutaneous Intended for the delivery of angioplasty balloons and other types of interventional devices.

DateAug 22, 2016 (initiated Apr 15, 2016)
StatusTerminated
ReasonIncrease in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.
Root causeUnknown/Undetermined by firm
ActionCook Inc. sent an URGENT: MEDICAL DEVICE RECALL NOTIFICATION letter dated April 15, 2016, to all affected customers. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers with any medical questions or concerns, were instructed to contact Cook Medical Customer Relations at 800.457.4500 or 812.339.2235. For information regarding the recall, please contact Stericycle Expert Solutions at 866.201.9067.
Quantity3,425
DistributionWorldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K902905COOK CORONARY GUIDING CATHETER
Cook, Inc., Sep 4, 1990