Recall Z-2841-2016 ยท CFL Radioimmunoassay, Human Growth Hormone

Siemens Healthcare Diagnostics, Inc.

IG1, IGF-I reagent, Catalog # LKGF1; For the quantitative measurement of insulin-like growth factor I (IGF-I) in serum or heparinized plasma.

DateSep 16, 2016 (initiated Apr 15, 2016)
StatusTerminated
Reason24 minutes of incubation is required prior to processing patient samples and for the pretreatment solution to reach full equilibration when using pretreatment solution when using the pretreatment solution (LGFA) contained in the kit.
Root causeUnknown/Undetermined by firm
ActionOn April 15, 2016, Siemens Healthcare Diagnostics Inc. distributed Urgent Medical Device Correction notices to their domestic customers and Urgent Field Safety Notices to their internationally customers via courier service. Customers are advised to take the following four actions: 1) To continue using IGF-I kit lot 411 and the associated pretreatment solution lot 055 to report patient results, ensure the IGF-I pretreatment solution is allowed to incubate with the patient sample for a minimum of 24 minutes prior to processing on the IMMULITE/IMMULITE 1000 systems. 2) Complete and return the E
Quantity525 units
DistributionDomestic: IL, FL, NY, OH, LA, MO, MA; International: Canada, others will be inserted