Recall Z-1884-2016 ยท HWE Instrument, Surgical, Orthopedic, Ac-Powered Motor And Accessory/Attachment
Greatbatch Medical
Torque Limiting Adapter: Torque Limiting Attachment, Non Sterile, Distributed by Zimmer. PRECIMED, Torque Limiting Adaptor, Non Sterile.. Customer Number/Model: 00-2360-080-00 T10232, T116245 T116245, TLA025-25 TLA025-25, 00-2360-080-05 T10233, MAC21003 T116430, TLA025-30 TLA025-30, MAC21001 T112972, MAC21001 T13621, TLA025-20 TLA025-20.
| Date | Apr 26, 2016 |
| Status | Terminated |
| Reason | Greatbatch Medical has initiated a global Field Recall for Torque Limiting Devices. Greatbatch identified through laboratory testing that the required sterility assurance level (SAL) of 10-6 is not achieved on torque limiting devices when subjected to steam sterilization cycles identified in the Instruction For Use (IFU) provided with the device. |
| Root cause | Device Design |
| Action | Consignees were sent on 4/26/2016 a Greatbatch "Urgent Medical Device Recall" letter dated April 26, 2016. The letter described the product involved in the recall, the reason, and the risk to health. The letter advised consignees to identity and quarantine the product and to complete and return the Filed Recall Response Form. For questions contact your local Greatbatch Medical representative, at 1-763-951-8235, or e-mail FieldActionCenter@Greatbatch.com. |
| Quantity | 5952 |
| Distribution | US: IN, MI, TX. OUS: GERMANY, FRANCE, ITALY, SWITZERLAND. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.