Recall Z-1720-2016 ยท JPA System, Multipurpose For In Vitro Coagulation Studies

Accriva Diagnostics Inc., dba ITC, dba Accumetrics

Accriva- APTT Cuvette for use on the Hemochron Jr. test system, catalog #J103, lot B6JCA012.

DateApr 22, 2016
StatusTerminated
ReasonClot activating reagent mis-positioned or visibly absent in the cuvette's test channel causing APTT test to either have an error code or an erroneously high result.
Root causeProcess design
ActionAccriva sent an " Urgent Field Safety Notice" letter dated April 22, 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were requested to take the following actions: 1. Forward communication to all those within organization who need to be aware of this matter. In addition, forward notification to all other organizations where affected devices may have been transferred to. 2. Check inventory to determine if you have any APTT Test Cuvettes, Lot Number: B6JCA012 3. If have inventory of APTT Test Cuvet
Quantity135 boxes
DistributionU.S. distribution to the following; state: VA. Foreign distribution to the following; China, Italy, Germany.