Recall Z-2078-2016 ยท FGB Ureteroscope And Accessories, Flexible/Rigid
Pentax Medical Company
Ureteroscope Intended to provide optical visualization of and therapeutic access to the Urinary Track. This anatomy includes but is not restricted to the organs, tissues and subsystems: Urethra, Bladder and Renal Pelvis. The instrument is introduced via the urethra when indications consistent with the requirement for procedure are observed in adult and pediatric patient populations.
| Date | Jun 28, 2016 (initiated Apr 6, 2016) |
| Status | Terminated |
| Reason | Pentax has become aware that operational/cleaning accessories and therapeutic devices (e.g., clips, stents, balloons, etc.) can become lodged in an endoscope's instrument channel. |
| Root cause | Device Design |
| Action | Pentax sent an Urgent Field Correction letter dated April 6, 2016 to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were reminded to carefully check that all accessories are intact, that no parts have fallen off and become lodged within the endoscope's instrument/suction channel and to ensure that any therapeutic devices passed through the instrument channel are accounted for after use. If the instrument/suction channel does become blocked or clogged due to the accumulation of debris, an accessory that cannot |
| Quantity | 32 units |
| Distribution | Worldwide Distribution - US (Nation Wide) including Puerto Rico and Internationally to Denmark, Guam |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K914106 | PENTAX URETERORENOFIBERSCOPE, MODEL FUR-10P Pentax Precision Instrument Corp., Jan 10, 1992 |