Recall Z-1793-2016 ยท HBA Neurosurgical Paddie

American Surgical Company

American Surgical Delicot X 9 mm x 50 mm Ref Number: 63x07 Product Usage: Neurosurgical and Nonneurosurgical Sponges

DateMar 31, 2016
StatusTerminated
ReasonProduct packaging defective compromising sterility
Root causePackaging process control
ActionAmerican Surgical issued an Urgent Recall and Field Safety Notice letter dated March 31, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were provided instruction to address affected devices in their inventory. They should address this by either selecting Action A or B: Action A: Users should perform a check of the device before use and, as stated in the IFU, patties should not be used if in an open or damaged pouch. For the affected lots, special attention should be given to lower portion of the pouch where the Tyvek and
Quantity1 box
DistributionWorldwide Distribution - US Nationwide and the countries of Australia, Belgium, Canada, France, Germany, Greece, Hong Kong, Hong Kong, Israel, Italy, Italy, Japan, Netherlands, New Zealand, Portugal, Saudi Arabia, South Africa, Spain, Sweden, Switzerland, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121822DELICOT
American Surgical Company, LLC, Sep 10, 2012