Recall Z-1560-2016 ยท JAK System, X-Ray, Tomography, Computed

Toshiba American Medical Systems Inc

Aquilion Lightning CT System iStation cover, SID: 70000246, Item Code: TSX-0354A/2D.03 Product Usage: Aquilion Lightning is a multislice helical CT system that supports whole-body imaging. The system employs Toshibas innovative dose-reduction technologies and a fast reconstruction unit designed to minimize the patient exposure dose and improve the workflow. The wide range of advanced applications

DateApr 12, 2016
StatusTerminated
ReasonToshiba American Medical Systems, Inc. is recalling the TSX-035A because of a potential problem with the Aquilion Lightning CT System iStation cover falling off of the gantry.
Root causeDevice Design
ActionAn Urgent Medical Device Correction letter dated 4/12/16 was sent to a customer to bring to the customer's attention with a potential problem with the Aquilion Lightning CT System iStation cover. The letter informs the customer that the iStation cover may come off due to deficient dimensional geomtry. The letter informs the customer of the problem identified and corrective action to be taken. The customer is instructed to complete and return the attached form and fax it to (877) 349-3054. If the customer has any questions regarding the recall letter, then the customer is instructed to cont
Quantity1 unit
DistributionUS in the state Nevada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151833Aquilion Lightning (TSX-035A/2, V7.0)
Toshibamedical Systems Corporation, Oct 9, 2015