ClearanceBrief
Recall Z-1634-2016 ยท MBI Fastener, Fixation, Nondegradable, Soft Tissue

Stryker Corporation

CinchLock SS Knotless Anchor with Inserter; Model number CAT02462; Orthopedic: The Pivot CinchLock SS Knotless Anchor (CAT02462) and Pivot CinchLock Flex Knotless Anchor (CAT02463) are non-degradable implant devices intended for the fixation of soft tissue to bone. Used for repair of hip, shoulder, foot and ankle, elbow, wrist and hand and knee.

DateApr 12, 2016
StatusTerminated
ReasonStryker Endoscopy is recalling the CinchLock SS Knotless Anchor with Inserter and the CinchLock Flex Anchor with Inserter after 13 reports of s of an internal wire breaking upon deployment of the anchor.
Root causeDevice Design
ActionUrgent Medical Device Recalls letters were sent on April 12, 2016 to US customers by certified mail. Stryker International distribution sites will be notified by e-mail. Letters describe the reason for recall with the affected part and lot numbers. Any recalled product should be segregated and returned to Stryker.
Quantity3100 total device, both model numbers.
DistributionDistributed US (nationwide) and in Australia, Switzerland, Netherlands, Spain and Sweden.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131769CINCHLOCK KNOTLESS SUTURE ANCHOR, NANOTACK SUTURE ANCHOR 1.4MM
Pivot Medical, Inc., Oct 7, 2013