Recall Z-1593-2016 ยท LLZ System, Image Processing, Radiological

Carestream Health Inc

CARESTREAM Image Suite V3: MINI-PACS/F IMG ST/CLASSIC: REF/Catalog # 1036490; MINI-PACS/F IMG ST/POC&VITA: REF/Catalog # 1036508; MINI-PACS/F IMG ST/CLASSIC/INDIA: REF/Catalog # 1036417; MINI-PACS/F IMG ST/ POC&VITA /INDIA: REF/Catalog # 1036425 -- Made in USA by: Carestream Health, Inc. 150 Verona Street, Rochester, NY 14608 --- CLASSIFICATION NAME: System, Image Processing, Radiological The C

DateMay 2, 2016 (initiated Apr 15, 2016)
StatusTerminated
ReasonCarestream Health received a complaint related to CARESTREAM Image Suite 4 from a foreign hospital stating that the annotation on the overlay is displayed as "L (Left)", when it should be "R(Right)".
Root causeSoftware design
ActionCarestream Health sent an URGENT MEDICAL DEVICE CORRECTION letter dated April 6, 2016, and Consignee Notification Acknowledgement form to the US consignees on April 15, 2016, via Certified Mail, Return Receipt Requested. Foreign consignees are being notified based on the individual country regulations. Customers who have this Image Suite configuration but do not use it for CT or MR Imaging do not need to take any action. Customers who use the product for CT or MR Images are asked to make users aware of the problem and take additional steps to check correct orientation of any markers, until t
QuantityUS: 50 units; Foreign: 2499 units
DistributionNY and OH for further distribution nationwide; Foreign distribution to the following countries: Afghanistan, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia & Herzgovena, Brazil, Bulgaria, Cambodia, Canada, Chad, Chile, China, Colombia, Croatia, Czech Rep., Djibouti, Fran
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K100094CARESTREAM IMAGE SUITE
Carestream Heatlh, Inc., Mar 11, 2010