Recall Z-1505-2016 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

Extended Brilliance Workspace NM Special Nuclear medicine image display and processing application suite. AutoSPECT Pro Software Application. Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.

DateMar 4, 2016
StatusTerminated
ReasonThe AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90 or 180 relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.
Root causeSoftware design
ActionOn March 4, 2016, the firm sent an URGENT - Field Safety Notice Medical Device Correction letter to their customers. The letter identified the affected device, the reason for the recall, and under what circumstances the problem occurs. Customers are to contact Philips as described if they are unsure whether their gamma cameras supports the modes. The firm will implement the AutoSPECT Pro software upgrade and contact customers to schedule the appointment. If further assistance or support is needed, customers are to contact their local Philips representative.
Quantity37 Units
DistributionWorldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South A
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090403AUTOSPECT
Philips Medical Systems (Cleveland), Inc., Mar 31, 2009