Recall Z-1534-2016 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

B Braun Medical Inc

Dialog+ Hemodialysis machines: Dialog + 120 V DF+Adimea+WAN-BSL-BIC Product Code # 710200U Serial # 202136 and < 203837 The Dialog+ can be used for implementing and monitoring hemodialysis treatments for patients with acute or chronic kidney failure. The system can be used in hospital, health center and outpatient dialysis center settings when prescribed by a Physician. The following types of rena

DateMay 3, 2016 (initiated Apr 1, 2016)
StatusTerminated
ReasonPotential leakage of conductivity sensors of the Dialog Dialysis machines.
Root causeProcess control
ActionB. Braun mailed an Urgent Medical Device Correction notice to affected customers informing them of the Dialog+Hemodialysis System due to the potential for Bicarbonate and End conductivity cells (3456102A and 3456103A respectively) to exhibit cracks. Notice provided customers with actions for Qualified Service Technicians, Clinical Personnel Administering Dialysis treatment and adverse event reporting. All Dialog+Hemodialysis machines will be inspected at the customer's site by the customer Biomed, who has been trained and qualified by B. Braun Avitum. The inspection of each machine will be pe
Quantity90 units
DistributionNationwide Distribution to ND, OH, CA, VA, FL, LA, SC, IL, UT, AZ, KY, IA, WI, OK, TN, SD, AL, NJ, TX, NM, NY, KS, WA, MO, CO, PA, CT, DE, GA, MD, VA, Puerto Rico, and St. Thomas USVI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083460DIALOG WITH ADIMEA OPTION
B. Braun Avitum AG, Feb 18, 2010
K963440DIALOG DIALYSIS MACHINE
B. Braun Medical AG, Jun 26, 1998