Recall Z-1532-2016 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System
B Braun Medical Inc
Dialog+ Hemodialysis machines: Dialog A + 1 Blood Pump 120V Product Code 710200K Serial # >202169 and < 203249 The Dialog+ can be used for implementing and monitoring hemodialysis treatments for patients with acute or chronic kidney failure. The system can be used in hospital, health center and outpatient dialysis center settings when prescribed by a Physician. The following types of renal replace
| Date | May 3, 2016 (initiated Apr 1, 2016) |
| Status | Terminated |
| Reason | Potential leakage of conductivity sensors of the Dialog Dialysis machines. |
| Root cause | Process control |
| Action | B. Braun mailed an Urgent Medical Device Correction notice to affected customers informing them of the Dialog+Hemodialysis System due to the potential for Bicarbonate and End conductivity cells (3456102A and 3456103A respectively) to exhibit cracks. Notice provided customers with actions for Qualified Service Technicians, Clinical Personnel Administering Dialysis treatment and adverse event reporting. All Dialog+Hemodialysis machines will be inspected at the customer's site by the customer Biomed, who has been trained and qualified by B. Braun Avitum. The inspection of each machine will be pe |
| Quantity | 40 units |
| Distribution | Nationwide Distribution to ND, OH, CA, VA, FL, LA, SC, IL, UT, AZ, KY, IA, WI, OK, TN, SD, AL, NJ, TX, NM, NY, KS, WA, MO, CO, PA, CT, DE, GA, MD, VA, Puerto Rico, and St. Thomas USVI |
| FDA record | Open on fda.gov |