Recall Z-1573-2016 ยท JAK System, X-Ray, Tomography, Computed

Philips and Neusoft Medical Systems Co., Ltd.

NeuViz 16 Multi-Slice CT Scanner System PN: 989605858501 a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.

DateApr 27, 2016 (initiated Mar 24, 2016)
StatusTerminated
ReasonThe following issues are found in NeuViz 16 systems with software version 1.1.4.21425 and version 1.1.4.21426: 1) During the filming operation on MX 16-slice console software, the clipboard used for copying and pasting images is not cleared between patients. If the operator fails to copy the current patient's images before pasting, a previous patient's image may be present in the clipboard and be
Root causeUnder Investigation by firm
ActionAffected consignees will be notified of the recall via letter on March 24, 2016. Mandatory Field Change Order and FCO kits will be released to all affected systems in June, 2016.
Quantity37 units
DistributionNC, OH, NE, SC, TX, LA, PR, MO, FL, CT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K090173NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM
Neusoft Medical Systems Co., Ltd., Feb 5, 2009
K092742NEUVIZ 16 MULTI-SLICE CT SCANNER SYSTEM
Neusoft Medical Systems Co., Ltd., Sep 22, 2009