ClearanceBrief
Recall Z-1830-2016 ยท OWD Somatic Gene Mutation Detection System

Roche Molecular Systems, Inc.

cobas EGFR Mutation Test, v2 and cobas cfDNA Sample Preparation Hungarian Translation Instructions for Use

DateMar 15, 2016
StatusTerminated
ReasonAn error was found within the Hungarian translations of the cobas EGFR Mutation Test v2 Instructions for Use (M/N 07340761001-01HU, Doc Rev. 1.0, Dated 08/2015) and the cobas cfDNA Sample Preparation Kit Instructions for Use (M/N 07573758001-01HU, Doc. Rev. 1.0, Dated 05/2015).
Root causePackaging
ActionRoche Molecular Diagnostics Inc. issued an Urgent Field Safety Notice dated March 15, 2016 to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Actions to be taken by the customer: " If you have utilized 2 L of DNA PBB during testing of plasma specimens, the samples must be retested using the correct volume (2mL DNA PBB). Communication of this Field Safety Corrective Action This notice must be passed on to all those who need to be aware within your organization or to any organization/individual where the potentially affected
Quantity8 kits
DistributionHungary
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.