ClearanceBrief
Recall Z-1992-2016 ยท JAI Couch, Radiation Therapy, Powered

Med Tec Inc

Protura Robotic Couch System Software, versions 1.6.1 (Part Number MT6XSM1.6.1), 1.7.0 (Part Number MT6XSM1.7.0), and 1.7.1 (Part Number MT6XSM1.7.1), CIVCO, Orange City, IA 51041

DateMar 31, 2016
StatusTerminated
ReasonWhen using the recalled product in split shift mode, the non-sub-millimeter positioning has been found to be inconsistent with the setting parameters.
Root causeSoftware design
ActionCivco Medical Solutions sent a "Urgent Field Correction Notice" dated March 31, 2016, to all affected customers. The letter identified the product the problem and the action need to be taken by the customer. The software's other configuration, "All 6 Degrees Protura" does not have the same translational issues. The firm send notification letters on February 23, 2016. In this letter, the Product Manager verified each site's software is configured to this operational mode which operates without the identified problem. In a later and separate notification on March 31, 2016, the firm informed
Quantity9
DistributionWorldwide Distribution - US Distribution including the countries of Switzerland, Germany..
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K132084PROTURA COUCH SOFTWARE
Med-Tec, Inc. D/B/A Civco Medical Solutions, Sep 26, 2013