Recall Z-1578-2016 ยท JJH Clinical Sample Concentrator

Roche Molecular Systems, Inc.

Hand-Held Scanner USB IT3800 For sample identification and tracking when used with various systems

DateApr 27, 2016 (initiated Mar 3, 2016)
StatusTerminated
ReasonThe hand-held barcode scanner model IT3800 used with the COBAS AmpliPrep instrument mis-identified a sample barcode ID.
Root causeLabeling design
ActionRoche Molecular Systems sent an "Urgent Medical Device Correction (UMDC) notification letter dated March 3, 2016, to their affected customers. Customers were instructed to provide the UMDC to other affected parties who may need to be aware of this issue. Customers were asked to complete the attached fax form and fax it to 1-866-737-5396. Customers with questions should contact Roche Support Network Customer Support Center at 1-800-526-1247. For questions regarding this recall call 908-253-7200.
Quantity6,939 pieces
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.