ClearanceBrief
Recall Z-1564-2016 ยท DXX System, Catheter Control, Steerable

Hansen Medical Inc

Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and away from the x-ray radiation source. The Accessory Kits contain disposable components that keep the Ma

DateMar 25, 2016
StatusTerminated
ReasonFirm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.
Root causeUnder Investigation by firm
ActionHansen Medical sent an Urgent Notice Product Recall notification dated March 28, 2016, to all affected customers. The notification identified the product, the problem, and the action to be taken by the customer. Customers were instructed to provide the product to the Hansen Medical field representative to assure credit appropriately applied to their account. Customers with questions were instructed to contact their local Hansen Medical representative or call 650-404-5581.
Quantity8 kits
DistributionNationwide Distribution to OH, TX. FL, CA and GA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111004HANSEN MEDICAL VASCULAR CATHETER CONTROL SYSTEM AND CATHETER
Hansen Medical, Inc., May 29, 2012
K132369MAGELLAN ROBOTIC SYSTEM, MAGELLAN ROBOTIC CATHETER 9FR
Hansen Medical, Inc., Sep 27, 2013
K141614MAGELLAN ROBOTIC SYSTEM
Hansen Medical, Aug 7, 2014
K151730Magellan Robotic System
Hansen Medical, Inc., Jul 23, 2015