Recall Z-1649-2016 ยท LZH Pump, Infusion, Enteral

Medtronic

Covidien Kangaroo Connect Enteral Feeding Pump Item Number: 584400 (OUS) Intended to provide enteral nutrition to a patient for hospital and acute care settings.

DateApr 1, 2016
StatusTerminated
ReasonKangaroo Connect Feeding Pump Occlusion alarms fail to alarm
Root causeDevice Design
ActionMedtronic(formerly) Covidien notified accounts via Urgent Medical Device Recall letter, dated 4/1/16, that specific serial numbers with software versions 1.12 and 1.13 have a software anomaly. The firm shipped replacement units with software V 1.14. Affected devices must be returned to the firm once the replacement units have been received. A Recalled Product Return Form required to be completed and returned. Questions or concerns, contact Medtronic representative or Customer Service at (800) 882-5878.
Quantity145 units
DistributionWorldwide Distribution -- USA, Australia, Canada, France, and Singapore.