ClearanceBrief
Recall Z-1845-2016 ยท FDT Duodenoscope And Accessories, Flexible/Rigid

Olympus Corporation of the Americas

EVIS EXERA Duodenovideoscope Olympus TJF Type 160F/VF used with an Olympus video system center, light source, documentation equipment monitors, Endo Therapy accessories and other ancillary equipment for endoscopy and endoscopic surgery within the duodenum.

DateMar 15, 2016
StatusTerminated
ReasonNew reprocessing instructions for the Olympus TJF-160/VF duodenoscopes, consisting of revised manual cleaning and high level disinfection procedures they plan to implement as well as a new cleaning brush (MAJ-1534), which is enclosed with the notification.
Root causeUnder Investigation by firm
ActionOlympus notified customers on 03/14/2016 via an Urgent Safety Notification letter that referenced discontinuation of previously distributed copies of the TJF-160F/VF Reprocessing Manual. Customers were asked to inspect their inventory and discard any existing inventory of TJF-160F/VF Reprocessing Manuals. They were asked to update their information with the enclosed TJF-160F/VF Reprocessing Manual, which contains new brushing and flushing steps for the TJF-160F/VF elevator mechanisms and the forceps elevator recesses. Instuctions for cleaning and a new cleaning brush (used for cleaning th
DistributionNationwide:
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K024033XTJF-160AF DUODENOVIDEOSCOPE
Olympus Optical Co., Ltd., Dec 20, 2002