Recall Z-1375-2016 ยท DRC Trocar

Merit Medical Systems, Inc.

Boston Scientific TSX Transseptal Needle, Model Number M004TSX60o; Used to create the primary puncture in the interatrial septum when passing an intruder and/or catheter through the septum from the right side of the heart to the left side.

DateMar 15, 2016
StatusTerminated
ReasonMerit Medical Systems, Inc. is voluntarily conducting a recall of one lot of the TSX" Transseptal Needle due to a discrepancy between the labeling and the actual configuration of the device in the packaging. The curve dimensional values on the labels of this lot incorrectly indicate a 50 curve, rather than the correct 86 curve
Root causeSoftware design
ActionThe consignee was provided verbal and written notification of the field action on March 15, 2016. There was a total of one (1) consignee, an OEM customer in the USA, who received 25 units. Written notification included product identification information, instruction to immediately quarantine any devices and discontinue distribution, and instructions to initiate a product recall in accordance with their company's established policies for customers to whom they shipped the affected Merit product. An email notification was sent to the Merit sales representative of the affected account. The sales
Quantity25
DistributionDistributed to one customer in the US.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011727TRANSSEPTAL NEEDLE/TROCAR
Thomas Medical Products, Inc., May 2, 2002