ClearanceBrief
Recall Z-1303-2016 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

Siemens branded SOMATOM Emotion and SOMATOM Sensation families of CT systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes taken at different angles.

DateFeb 15, 2016
StatusTerminated
ReasonSiemens is initiating this recall due to a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems in conjunction with the use of the cardio licenses "Cardio Pro" and "Cardio Retro".
Root causeDevice Design
ActionA Safety Advisory Notice was sent to customers informing them of a potential malfunction and hence potential hazard to patients and personnel when using the SOMATOM Emotion or SOMATOM Sensation systems. Siemens informed them that they will deliver an Update Instruction for their Service Engineers to modify the installed Firmware. This will correct the rare synchronization errors of Scanner Firmware and Scanner Software to start the application of the x-ray in compliance with the optimum triggering point for best image quality.
Quantity124 units
DistributionUS (nationwide) including Guam and the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, KY, MA, MD, ME, MI, MN, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, VA, and WA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K013522SOMATOM P30 CT SYSTEMS
Siemens Medical Systems, Inc., Nov 7, 2001